Chinese Journal of Radiation Oncology ; (6): 89-93, 2018.
Article in Chinese | WPRIM | ID: wpr-666095

ABSTRACT

Objective To analyze the precision and stability of optical surface imaging system for patients who received radiotherapy with active breath control. Methods Eighteen radiotherapy patients with lung metastasis were managed by active breath control (ABC).The difference error detected by optical surface imaging system and CBCT were defined as the precision of optical surface imaging system. The variation among the error of optical surface imaging positioning the value of correction of treatment position and the error detected by optical surface imaging again were defined as the stability of optical surface imaging system. Intrafractional errors were analyzed by optical surface imaging system through whole treatment process (including breath hold and free breath). Results The optical surface imaging system had precision (systematic (Σ) and random errors (σ)) of 1.78/3.42 mm 2.54/6.57 mm and 2.79/3.22 mm respectively and stability of2.12/2.54 mm 3. 09/4.02 mm and 1.37/3.55 mm respectively in lateral-medial superior-inferior and anterior-posterior directions. The intrafractional errors (Σ and σ) were 0.42/0.85 mm 0.41/1.47 mm and 0.41/1.47 mm respectively for breath hold duration and 4.76/4.16 mm 6.54/7.78 mm and 3.13/5.92 mm for free breath duration in lateral-medial superior-inferior and anterior-posterior directions. Conclusions As an effective method for validate breath hold;Optical surface imaging system can improve the precision and safety of active breath control. However,the factors that affect the accuracy and stability of the optical surface imaging system in patients undergoing radiotherapy with ABC are not clear;it cannot replace the CBCT for positioning verification.

2.
Chinese Journal of Radiation Oncology ; (6): 768-773, 2017.
Article in Chinese | WPRIM | ID: wpr-620219

ABSTRACT

Objective To analyze the displacement of titanium clips for tumor bed localization after breast-conserving surgery for breast cancer and its influential factors.Methods A retrospective analysis was performed on the cone-beam computed tomography (CT) images of 14 patients with breast cancer who received radiotherapy after breast-conserving surgery from April to October,2016.The relative position of the chest wall and the errors of the titanium clips in radiotherapy were measured.A Pearson correlation analysis was used to analyze the correlation of the displacement of titanium clips with the relative position of titanium clips,the breast volume,the vertical distance between the titanium clips and the tangential line of the chest wall,and the maximum thickness of the breast.Results The system errors of the chest wall in left-right,superior-inferior,and anterior-posterior directions were 4.42,3.44,and 5.13 mm,respectively,and the random errors were 3.55,3.07,and 4.54 mm,respectively.The titanium clips had a large displacement relative to the chest wall,mainly in the left-right direction.The maximum system error was 4.39 mm and the random error was 2.42 mm.The displacement of titanium clips was not significantly correlated with the breast volume and the maximum thickness of the breast (P>0.05).However,the relative position of titanium clips in superior-inferior direction was significantly correlated with the displacement of the lowest,the most lateral,the most anterior,and the most posterior titanium clips (P<0.05).As to the uppermost clips,there was a significant difference in displacement between the clips close to the chest wall and the clips far from the chest wall (P=0.02).Conclusions Due to large setup error and displacement of titanium clips during radiotherapy,simultaneous integrated boost is not suitable for patients with breast cancer who are immobilized by vacuum cushion and received radiotherapy.The unstable immobilization may be the major influential factor for the displacement of titanium clips.

SELECTION OF CITATIONS
SEARCH DETAIL